ELISA standard curve calculator (four-parameter logistic, 4PL)
Excel has no 4PL trendline, and building one with Solver takes time. Paste the concentrations and ODs of your standards and the ODs of your samples, and this page fits the S-shaped curve and gives the sample concentrations. Samples outside the range of the standards are flagged rather than extrapolated.
Your numbers are processed only in this browser. Nothing is sent to a server (it is gone when you close the page).
One row = one concentration (or one sample). Replicate wells can be side by side or on repeated rows with the same concentration or name. A sideways table (concentrations across the first row, as plate readers export it) is turned upright automatically. A header row is optional; a unit in brackets such as “(pg/mL)” is used in the chart and tables. Sample dilution factors are read from a column headed “Dilution”, “DF” or “Factor”.
Used for samples without a dilution column. Concentration × dilution is the concentration in the original sample.
Table layout
Paste the standards on the left and the samples on the right. Select the range in Excel, copy, and paste.
| Conc (pg/mL) | OD 1 | OD 2 |
|---|---|---|
| 500 | 2.195 | 2.102 |
| 250 | 1.570 | 1.561 |
| … | … | … |
| 7.8 | 0.117 | 0.127 |
| 0 | 0.054 | 0.058 |
| Sample | OD 1 | OD 2 | Dilution |
|---|---|---|---|
| Serum 1 | 0.452 | 0.468 | 2 |
| Serum 2 | 1.105 | 1.142 | 2 |
| Supernatant | 1.650 | 1.598 | 1 |
- A label column such as “S1, S2…” before the concentrations is fine (the second column is then read as concentration).
- Replicate wells can be side by side or on repeated rows, and the number of wells may differ between rows.
- A sample row with an empty name is read as the sample above (merged cells in Excel).
- Blank and N/A cells are left out. Decimal commas (0,125) are understood.
The 4PL equation and its four parameters
This page uses the following equation (x = concentration, y = OD).
y = d + (a − d) / (1 + (x / c)b)
| Symbol | Meaning | Example data |
|---|---|---|
| a | OD at zero concentration (lower plateau; upper in a competitive assay) | 0.05810 |
| d | OD at very high concentration (upper plateau; lower in a competitive assay) | 3.240 |
| c | EC50 (inflection point): the concentration giving an OD halfway between a and d; on a log concentration axis, this is where the curve is steepest | 276.2 pg/mL |
| b | Slope (Hill slope): larger means a steeper S | 1.097 |
The fit uses concentration itself (fitting on log concentration gives the same curve, with c replaced by log10 c). At zero concentration (x/c)b = 0, so y = a and the zero standard can be used in the fit as it is. The curve is fitted by the Levenberg–Marquardt method (nonlinear least squares) from several starting values, keeping the best fit.
Why 4PL rather than a straight line
ELISA ODs level off at both low and high concentrations, giving an S-shaped curve. A straight line (or a log–log line) fits the middle but drifts at both ends, so low and high samples come out wrong. This is why S-shaped models such as 4PL are widely used for ELISA standard curves (follow your kit manual if it specifies a model).
Excel’s trendlines do not include 4PL, so people usually build a Solver sheet to find a, b, c and d. This page saves that step. When comparing with other software, check that the equation, the weighting and the handling of replicate wells (each well as a point, or means) are the same; if any of them differ, the values will differ slightly.
Handling the zero standard (blank)
- Include in the fit (default): zero-concentration wells are used as one more standard. The 4PL parameter a is then the zero-dose OD, so the curve is correct without blank subtraction. Use this for the B0 of a competitive ELISA too.
- Subtract as blank: the mean OD of the zero wells is subtracted from every standard and sample OD before fitting (the zero wells are then left out of the fit). Use this when the kit protocol says to subtract the blank.
Concentrations of in-range samples usually change very little between the two. Follow the kit protocol and state which one you used.
Choosing the weighting (none or 1/Y²)
OD variability often grows with the OD. Without weighting, the noisy high-OD points pull the curve and the low-concentration end can suffer. 1/Y² weighting equalises the relative (percentage) error and gives more weight to low ODs.
- Start with no weighting and look at the %recovery of the standards (table below)
- If low standards recover outside your acceptance range (e.g. 80–120%), or low samples matter most, try 1/Y²
- 1/Y² cannot be used when an OD is zero or negative (common after blank subtraction; include the zero standard instead)
- State which you used (the methods sentence below includes it automatically)
On this page, 1/Y² uses the measured (observed) OD, as when the OD is passed as sigma to SciPy’s curve_fit. Prism (version 5 and later) weights by the fitted curve’s value rather than the measured OD, so with 1/Y² the results can differ slightly (without weighting, the same equation gives the same values).
Samples outside the range (why no extrapolation)
Samples outside the range of the standards (from the lowest non-zero to the highest standard) are reported as “below range” or “above range”, without a concentration, for two reasons.
- Outside the range, the curve has not been checked against standards. Extending it (extrapolation) can be far off.
- Near the plateaus the 4PL curve flattens, so a tiny OD difference means a large concentration difference. ODs beyond a plateau cannot be converted at all.
What to do: if a sample is above range, dilute it further and measure again. If it is below range, report it as “< lowest standard × dilution” (e.g. < 15.6 pg/mL), following the rules of your field or journal for values below the lowest standard (the lower limit of the quantification range). The tables and the copied results use this form.
Judging the curve (not by R² alone)
With four parameters, a 4PL curve is flexible and R² is usually above 0.99, even when the low end is off. Look at these as well.
| Check | Rule of thumb (varies by lab and guideline) | On this page |
|---|---|---|
| %Recovery of standards | back-calculated ÷ nominal × 100 within 80–120% | shown in red if outside |
| CV between wells | CV of replicate wells (SD ÷ mean × 100) ≤ 15% | shown in red if above 15% |
| Number of standards | 6–8 non-zero concentrations, 2+ wells each | warning below 5 |
If only one end standard fails, you may drop it and narrow the range (record why). A high CV points to pipetting or washing problems. For how to count replicates, see biological vs technical replicates (wells of the same sample are technical replicates).
How it was checked
The parameters a, b, c and d with their standard errors and 95% CIs, R², back-calculated standards and %recovery, and sample concentrations match an independent Python calculation (SciPy 1.18 curve_fit, same equation and weighting) to a relative error below 10−5 for the parameters. The checks cover the example data (zero standard included or subtracted × no weighting or 1/Y²), a decreasing competitive ELISA, a noisy data set (triplicates, upper plateau poorly defined) and a table with a label column, decimal commas and a missing well.
For the example (no weighting, zero standard included): a = 0.05810, b = 1.097, c = 276.2 pg/mL, d = 3.240, R² = 0.9992; Serum 1 is 47.41 pg/mL × 2 = 94.82 pg/mL. With 1/Y², c = 290.3 pg/mL and Serum 1 is 94.72 pg/mL. Standard errors are the usual asymptotic ones from the Jacobian (as in curve_fit). The 95% CI of c is a Wald interval on the log scale (c × exp(± t × SE(c)/c)), as on the IC50 page, so it never goes below 0 (a, b and d use the usual ± t × SE).
Only 4PL is available for now; 5PL (an asymmetric S-curve) is not yet supported.
FAQ
Is my data sent anywhere?
No. Everything runs in your browser and the ODs and concentrations are never sent to a server. They disappear when you close the page.
Can I fit a 4PL curve in Excel?
Excel’s trendlines do not include 4PL, so you would have to build a Solver sheet to find a, b, c and d. Here you just paste the data, and “Copy all tables” gives a table you can paste back into Excel.
What R² is good enough?
A 4PL fit usually gives R² above 0.99, so R² alone says little. Check the %recovery of the standards and the CV of replicate wells too. Common rules of thumb are 80–120% recovery and CV ≤ 15%, but limits vary by lab and guideline.
How do I get a value for an out-of-range sample?
If it is above range, dilute it more and measure again. If it is below range, report “< lowest standard × dilution”. Extrapolating the curve can be badly wrong, so this page does not give a value.
Does it work for a competitive ELISA (OD falls as concentration rises)?
Yes. The direction of the curve is detected automatically (a > d means decreasing). Keep the zero standard (B0) as “Include in the fit”. You can also paste B/B0 (%) instead of OD.
Will I get the same values as other software?
Values are close only when the equation (how the 4PL is written), the weighting, the data points (each replicate well as its own point, or means) and the handling of the zero standard all match. This page uses each replicate well as a point, and 1/Y² weights by the measured OD. If any of these differ, expect small differences.