ELISA standard curve calculator (four-parameter logistic, 4PL)

Excel has no 4PL trendline, and building one with Solver takes time. Paste the concentrations and ODs of your standards and the ODs of your samples, and this page fits the S-shaped curve and gives the sample concentrations. Samples outside the range of the standards are flagged rather than extrapolated.

Updated: BenchFig日本語

Your numbers are processed only in this browser. Nothing is sent to a server (it is gone when you close the page).

One row = one concentration (or one sample). Replicate wells can be side by side or on repeated rows with the same concentration or name. A sideways table (concentrations across the first row, as plate readers export it) is turned upright automatically. A header row is optional; a unit in brackets such as “(pg/mL)” is used in the chart and tables. Sample dilution factors are read from a column headed “Dilution”, “DF” or “Factor”.

Zero standard
Weighting

Used for samples without a dilution column. Concentration × dilution is the concentration in the original sample.

Table layout

Paste the standards on the left and the samples on the right. Select the range in Excel, copy, and paste.

Conc (pg/mL)OD 1OD 2
5002.1952.102
2501.5701.561
………
7.80.1170.127
00.0540.058
SampleOD 1OD 2Dilution
Serum 10.4520.4682
Serum 21.1051.1422
Supernatant1.6501.5981

The 4PL equation and its four parameters

This page uses the following equation (x = concentration, y = OD).

y = d + (a − d) / (1 + (x / c)b)

SymbolMeaningExample data
aOD at zero concentration (lower plateau; upper in a competitive assay)0.05810
dOD at very high concentration (upper plateau; lower in a competitive assay)3.240
cEC50 (inflection point): the concentration giving an OD halfway between a and d; on a log concentration axis, this is where the curve is steepest276.2 pg/mL
bSlope (Hill slope): larger means a steeper S1.097
Kit manuals and software may use different letters. When the equation is written with A, B, C and D, which of A and D is the lower plateau, and whether B carries a sign, differs between programs. Compare by meaning (zero-dose OD, plateau OD, inflection point, slope), not by letter. This equation has the same form as GraphPad Prism’s “Sigmoidal, 4PL, X is concentration” (Y = Bottom + XHillSlope(Top − Bottom)/(XHillSlope + EC50HillSlope)), with Bottom = a, Top = d, EC50 = c and HillSlope = b (in Prism’s nonlinear regression dialog it is called “[Agonist] vs. response -- Variable slope”: GraphPad guide).

The fit uses concentration itself (fitting on log concentration gives the same curve, with c replaced by log10 c). At zero concentration (x/c)b = 0, so y = a and the zero standard can be used in the fit as it is. The curve is fitted by the Levenberg–Marquardt method (nonlinear least squares) from several starting values, keeping the best fit.

Why 4PL rather than a straight line

ELISA ODs level off at both low and high concentrations, giving an S-shaped curve. A straight line (or a log–log line) fits the middle but drifts at both ends, so low and high samples come out wrong. This is why S-shaped models such as 4PL are widely used for ELISA standard curves (follow your kit manual if it specifies a model).

Excel’s trendlines do not include 4PL, so people usually build a Solver sheet to find a, b, c and d. This page saves that step. When comparing with other software, check that the equation, the weighting and the handling of replicate wells (each well as a point, or means) are the same; if any of them differ, the values will differ slightly.

Handling the zero standard (blank)

Concentrations of in-range samples usually change very little between the two. Follow the kit protocol and state which one you used.

Choosing the weighting (none or 1/Y²)

OD variability often grows with the OD. Without weighting, the noisy high-OD points pull the curve and the low-concentration end can suffer. 1/Y² weighting equalises the relative (percentage) error and gives more weight to low ODs.

Key points
  • Start with no weighting and look at the %recovery of the standards (table below)
  • If low standards recover outside your acceptance range (e.g. 80–120%), or low samples matter most, try 1/Y²
  • 1/Y² cannot be used when an OD is zero or negative (common after blank subtraction; include the zero standard instead)
  • State which you used (the methods sentence below includes it automatically)

On this page, 1/Y² uses the measured (observed) OD, as when the OD is passed as sigma to SciPy’s curve_fit. Prism (version 5 and later) weights by the fitted curve’s value rather than the measured OD, so with 1/Y² the results can differ slightly (without weighting, the same equation gives the same values).

Samples outside the range (why no extrapolation)

Samples outside the range of the standards (from the lowest non-zero to the highest standard) are reported as “below range” or “above range”, without a concentration, for two reasons.

  1. Outside the range, the curve has not been checked against standards. Extending it (extrapolation) can be far off.
  2. Near the plateaus the 4PL curve flattens, so a tiny OD difference means a large concentration difference. ODs beyond a plateau cannot be converted at all.

What to do: if a sample is above range, dilute it further and measure again. If it is below range, report it as “< lowest standard × dilution” (e.g. < 15.6 pg/mL), following the rules of your field or journal for values below the lowest standard (the lower limit of the quantification range). The tables and the copied results use this form.

Judging the curve (not by R² alone)

With four parameters, a 4PL curve is flexible and R² is usually above 0.99, even when the low end is off. Look at these as well.

CheckRule of thumb (varies by lab and guideline)On this page
%Recovery of standardsback-calculated ÷ nominal × 100 within 80–120%shown in red if outside
CV between wellsCV of replicate wells (SD ÷ mean × 100) ≤ 15%shown in red if above 15%
Number of standards6–8 non-zero concentrations, 2+ wells eachwarning below 5

If only one end standard fails, you may drop it and narrow the range (record why). A high CV points to pipetting or washing problems. For how to count replicates, see biological vs technical replicates (wells of the same sample are technical replicates).

How it was checked

The parameters a, b, c and d with their standard errors and 95% CIs, R², back-calculated standards and %recovery, and sample concentrations match an independent Python calculation (SciPy 1.18 curve_fit, same equation and weighting) to a relative error below 10−5 for the parameters. The checks cover the example data (zero standard included or subtracted × no weighting or 1/Y²), a decreasing competitive ELISA, a noisy data set (triplicates, upper plateau poorly defined) and a table with a label column, decimal commas and a missing well.

For the example (no weighting, zero standard included): a = 0.05810, b = 1.097, c = 276.2 pg/mL, d = 3.240, R² = 0.9992; Serum 1 is 47.41 pg/mL × 2 = 94.82 pg/mL. With 1/Y², c = 290.3 pg/mL and Serum 1 is 94.72 pg/mL. Standard errors are the usual asymptotic ones from the Jacobian (as in curve_fit). The 95% CI of c is a Wald interval on the log scale (c × exp(± t × SE(c)/c)), as on the IC50 page, so it never goes below 0 (a, b and d use the usual ± t × SE).

Only 4PL is available for now; 5PL (an asymmetric S-curve) is not yet supported.

FAQ

Is my data sent anywhere?

No. Everything runs in your browser and the ODs and concentrations are never sent to a server. They disappear when you close the page.

Can I fit a 4PL curve in Excel?

Excel’s trendlines do not include 4PL, so you would have to build a Solver sheet to find a, b, c and d. Here you just paste the data, and “Copy all tables” gives a table you can paste back into Excel.

What R² is good enough?

A 4PL fit usually gives R² above 0.99, so R² alone says little. Check the %recovery of the standards and the CV of replicate wells too. Common rules of thumb are 80–120% recovery and CV ≤ 15%, but limits vary by lab and guideline.

How do I get a value for an out-of-range sample?

If it is above range, dilute it more and measure again. If it is below range, report “< lowest standard × dilution”. Extrapolating the curve can be badly wrong, so this page does not give a value.

Does it work for a competitive ELISA (OD falls as concentration rises)?

Yes. The direction of the curve is detected automatically (a > d means decreasing). Keep the zero standard (B0) as “Include in the fit”. You can also paste B/B0 (%) instead of OD.

Will I get the same values as other software?

Values are close only when the equation (how the 4PL is written), the weighting, the data points (each replicate well as its own point, or means) and the handling of the zero standard all match. This page uses each replicate well as a point, and 1/Y² weights by the measured OD. If any of these differ, expect small differences.