qPCR standard curve and amplification efficiency (with a free Excel template)

A qPCR standard curve plots Cq against log(quantity) for a dilution series of a known standard. The slope gives the amplification efficiency, and reading an unknown’s Cq off the line gives its quantity. This worked example goes from slope and R² to efficiency and unknown quantities, and comes with a free Excel template that does the same.

Updated: BenchFig日本語

What a standard curve is

With perfect efficiency, PCR doubles the product every cycle, so a 10-fold higher starting amount reaches the threshold about 3.32 cycles earlier (23.32 ≈ 10). Measure a dilution series of a known standard and plot Cq (y) against log10(quantity) (x): that line is the standard curve.

Key points
  • Efficiency from the slope: efficiency = 10−1/slope − 1 (slope −3.32 = 100%)
  • Common acceptance: efficiency 90–110% (slope about −3.58 to −3.10), R² ≥ 0.98
  • Unknown quantity = 10(Cq − intercept) / slope; outside the curve it is extrapolated
  • Five or more levels, triplicates, spanning the unknowns’ Cq values

Designing the dilution series

Worked example

Five 10-fold dilutions from 105 to 10 copies, three wells each (made-up data).

Copieslog10(quantity)Cq (3 wells)
100,000517.12, 17.25, 17.08
10,000420.51, 20.44, 20.63
1,000323.86, 23.95, 23.79
100227.18, 27.31, 27.24
10130.62, 30.49, 30.71
0123456161820222426283032log₁₀ (starting quantity)CqStd 1-1: Cq 17.12, log SQ 5.00Std 1-2: Cq 17.25, log SQ 5.00Std 1-3: Cq 17.08, log SQ 5.00Std 2-1: Cq 20.51, log SQ 4.00Std 2-2: Cq 20.44, log SQ 4.00Std 2-3: Cq 20.63, log SQ 4.00Std 3-1: Cq 23.86, log SQ 3.00Std 3-2: Cq 23.95, log SQ 3.00Std 3-3: Cq 23.79, log SQ 3.00Std 4-1: Cq 27.18, log SQ 2.00Std 4-2: Cq 27.31, log SQ 2.00Std 4-3: Cq 27.24, log SQ 2.00Std 5-1: Cq 30.62, log SQ 1.00Std 5-2: Cq 30.49, log SQ 1.00Std 5-3: Cq 30.71, log SQ 1.00y = -3.363x + 33.97R² = 0.9998E = 98.3 %
Example standard curve (five 10-fold dilutions × 3 wells, made-up data). x = log10(copies), y = Cq. Dashed line: regression on all 15 wells. Same drawing as BenchFig’s qPCR screen.

1. Slope, intercept and R²

Fit a least-squares line to all 15 (log10(quantity), Cq) pairs.

Cq = −3.363 × log10(quantity) + 33.968  R2 = 0.99976Slope, intercept and R² match SciPy (scipy.stats.linregress).

2. Amplification efficiency

Efficiency = 10−1/slope − 1 = 10−1/(−3.363) − 1 = 1.983 − 1 = 98.3%1.983 is the amplification factor per cycle; 2 means 100%.
Slope vs efficiency. A steeper slope means lower efficiency.
SlopeEfficiency
−3.10110.2%
−3.20105.4%
−3.32100.1%
−3.4096.8%
−3.5890.3%
−3.8083.3%

3. Quantifying unknowns

Quantity = 10(Cq − intercept) / slope = 10(22.42 − 33.968) / (−3.363) = 2,715 copies
Standards span Cq 17.08–30.71. Sample C is later than the most dilute standard (10 copies), so its value is extrapolated.
SampleCq (3 wells)Mean CqEstimated copiesWithin curve?
Sample A22.40, 22.51, 22.3522.422,715Yes
Sample B25.90, 26.02, 25.8825.93245Yes
Sample C31.85, 31.97, 31.7031.844.29No (approximate)

When efficiency is out of range

A widely used acceptance range is efficiency 90–110% with R² ≥ 0.98 (Taylor et al. 2010; BenchFig’s qPCR screen warns outside this range).

SymptomCommon causesWhat to check
Efficiency above 110%Inhibition in the most concentrated points (delayed Cq), primer-dimers or non-specific products, pipetting errorRefit without the top point; melt curve; NTC Cq
Efficiency below 90%Suboptimal primer design or concentration, annealing temperature, long ampliconOptimise primer concentration and temperature, or redesign
Low R²Dilution error, noise near the detection limit, replicate scatterReplicate SD per level; refit without the most dilute point
Curved pointsLevels outside the linear rangeUse only the linear dynamic range and report it
Standard curves straight from CFX exportsLoad the CFX “Quantification Cq Results” to get per-target curves, slope, R², efficiency, flagged wells, unknown quantities and a graph. Try it with the sample file.
Open for free →

In Excel (free template)

Put one well per row with columns for log10(quantity) and Cq, then use these functions (Excel 2010 or later).

QuantityExcel formula (column C = log10(quantity), D = Cq)
log10(quantity)=LOG10(B6)
Slope=SLOPE(D6:D20,C6:C20) (y range first)
Intercept=INTERCEPT(D6:D20,C6:C20)
R²=RSQ(D6:D20,C6:C20)
Efficiency (%)=(10^(-1/G6)-1)*100 (G6 = slope)
Unknown quantity=10^((E24-$G$7)/$G$6) (E24 = mean Cq, G7 = intercept)
Download the standard curve Excel template (.xlsx)Free, no sign-up. Slope, intercept, R², efficiency and a range check, plus unknown quantities with out-of-range flags. Japanese version here

The template’s sample values match this example (slope −3.363, R² 0.99976, efficiency 98.3%, Sample A about 2,715 copies). Excel calculates them when the file opens.

How to report it

The MIQE guidelines (Bustin et al. 2009) ask for the slope (or efficiency), y-intercept, R² and linear dynamic range of each standard curve.

Amplification efficiencies were determined from standard curves of five 10-fold dilutions (105 to 10 copies, triplicate wells; slope −3.363, intercept 33.97, R² = 0.9998, efficiency 98.3%).

To choose between ΔΔCt and efficiency-corrected Pfaffl quantification, see ΔΔCt vs Pfaffl vs standard curve; for the calculation, the ΔΔCt method; for CFX files, from CFX data to graph. Regression and R² are explained in linear regression and R².

Sources: Bustin SA, et al. (2009) The MIQE guidelines. Clin Chem 55:611–622; Taylor S, et al. (2010) A practical approach to RT-qPCR—publishing data that conform to the MIQE guidelines. Methods 50:S1–S5 (90–110% efficiency, R²); Pfaffl MW (2001) Nucleic Acids Res 29:e45 (efficiency-corrected relative quantification).

FAQ

My efficiency is above 100%. Is that wrong?

The theoretical maximum is 100% (doubling every cycle), so values above 110% mean something is distorting the curve: inhibition in the most concentrated points (delayed Cq, flatter slope), pipetting error, or primer-dimers and other non-specific products. Drop the most concentrated point and check linearity, then check the melt curve.

Should R² be calculated from all wells or replicate means?

Both are used, but all wells is more common. Means hide replicate scatter, so R² looks higher (here 0.9998 with all wells vs 1.0000 with means). The slope, and hence efficiency, is the same either way when every level has the same number of replicates.

How many dilution points do I need?

Five or more levels, each in triplicate, is common. Five 10-fold dilutions span four logs; 4- or 5-fold dilutions give denser points. Above all, unknowns should fall inside the curve.

What about samples outside the curve?

Extrapolated values are not reliable; treat them as approximate. Samples with Cq later than the most dilute standard are near the detection limit and noisy. Dilute or concentrate and rerun, or extend the curve.

Do I need a standard curve for relative quantification?

The ΔΔCt method assumes efficiencies near 100% and similar for target and reference, so measure efficiency with a standard curve at least once. If efficiencies differ, use the Pfaffl method (see ΔΔCt vs Pfaffl).

Open the sample standard curve in BenchFigThe qPCR screen opens with a sample file that includes standard curves. Load a CFX export to get slope, R², efficiency, unknown quantities, optional Pfaffl correction and a graph automatically.
Open for free →